Title:
COMPARISON BETWEEN TONOVET, TONOVET PLUS, TONO-PEN AVIA VET AND KOWA HA-2 HANDHELD PORTABLE TONOMETERS IN THE MEASUREMENT OF INTRAOCULAR PRESSURE IN DOGS
Type:
Free Communication
Topic:
Ophthalmology
Companies:
(1) Post Graduate Program in Animal Science, University of Oeste Paulista (UNOESTE), Presidente Prudente, SP, Brazil
(2) Faculty of Veterinary Medicine (UNOESTE)
(3) Department of Small Animal of Veterinary Hospital (UNOESTE)
Authors:
João Victor Goulart Consoni Passareli (1)
Danielle Alves Silva (1)
Zulim Luís Felipe Da Costa (1)
Felipe Franco Nascimento (1)
Giovana José Garcia Estanho (1)
Glaucia Prada Kanashiro (1)
Bruna Cristina Da Cruz2 (2)
Jéssica Aparecida De Souza (2)
Thais Cardoso Araújo (2)
Silvia Franco Andrade (3)
Paper:
RESUMEN CORTO - SHORT SUMMARY

OBJECTIVE: Compare the accuracy of the tonometers, Tonovet, Tonovet Plus, Tono-Pen Avia Vet, and Kowa HA-2 in dogs. METHODS: A postmortem study (n = 10) comparing the values ​​of manometry versus tonometry and calculating the correlation coefficient (r2). An in vivo study (n = 10) was performed comparing the real IOP (manometry) with the tonometers and an ambulatory study in anesthetized healthy dogs (n = 20). An ambulatory study, eyes with signs of glaucoma (n = 10) and uveitis (n = 10) were analyzed. RESULTS: The post-mortem study of (r2) in decreasing order: Kowa HA-2 (0.989), Tonovet Plus (0.984), Tonovet (0.981), and Tono-Pen (0.847). The IOP values ​​in mmHg in the in vivo study: aneroid manometer (16.8 ± 1.7), Tonovet (18.1 ± 2.9), Tonovet Plus (20.6 ± 2.3), Tone-Pen (17.1 ± 2.5), and Kowa (16.1 ± 1.7). In the ambulatory study: healthy Tonovet (16.2 ± 2.8), Tonovet Plus (19.9 ± 3.4), Tono-Pen (15.3 ± 3.0), and Kowa (14.6 ± 1.9); glaucoma Tonovet (30.9 ± 5.9), Tonovet Plus (35.0 ± 8.3), Tono-Pen (28.4 ± 5.8), and Kowa (23.8 ± 4.4); and uveitis Tonovet (14.2 ± 5.6), Tonovet Plus (17.9 ± 5.5), Tono-Pen (13.3 ± 6.3) and Kowa (10.7 ± 4.0). CONCLUSIONS: There was a strong correlation between the IOP values with the manometry and all tonometers. The highest values were mensuremented with Tonovet Plus and the lowest with Kowa. All tonometers were accurate in the IOP measurement in dogs, including the most recent, Tonovet Plus, which showed an excellent correlation coefficient.



BIBLIOGRAFÍA

BIBLIOGRAPHY

1. Okafor KC, Brandt JD. Measuring intraocular pressure. Current Opinion in Ophthalmology 2015; 26(2):103-119.

2. Featherstone HJ, Heinrich CL. Ophthalmic examination and diagnostic. In: Gelatt KN, Gilger, BC, Kern TJ (ed). Veterinary Ophthalmology, Iowa, Willey-Blackwell, 2013; 533-613.

3. Leiva M, Naranjo C, Peña MT. Comparision of the rebound tonometer (ICare) to the applanation tonometer (Tonopen XL) in normotensive dogs. Vet Ophthalmol 2006; 9(1):17-21.

4. Andrade SF, Andrade AL, Kanashiro GP et al. Ex vivo and in vivo study of Kowa HA-2 applanation tonometer in the measurement of intraocular pressure in dogs. Semina 2016; 37(2): 797-806.

5. von Spiessen L1, Karck J, Rohn K, Meyer-Lindenberg A. Clinical comparison of the TonoVet(®) rebound tonometer and the Tono-Pen Vet(®) applanation tonometer in dogs and cats with ocular disease: glaucoma or corneal pathology. Vet Ophthalmol 2015;18(1):20-27.

6. Andrade SF, Palozzi RJ, Giuffrida R. Comparison of intraocular pressure measurements between the Tono-Pen XL® and Perkins® applanation tonometers in dogs and cats. Vet Ophthalmol 2012; 15(Supplement 1): 14–20.

7. Grewal DS, Stinnett SS, Folgar FA et al. A Comparative study of rebound tonometry with Tonopen and Goldmann applanation tonometry following vitreoretinal surgery. Am J Ophthalmol 2016; 161:22-28.